Gas analyzer range verification and drift verification.

§ 1065.550 Gas analyzer range verification and drift verification.

(a) Range verification. If an analyzer operated above 100% of its range at any time during the test, perform the following steps:

(1) For batch sampling, re-analyze the sample using the lowest analyzer range that results in a maximum instrument response below 100%. Report the result from the lowest range from which the analyzer operates below 100% of its range.

(2) For continuous sampling, repeat the entire test using the next higher analyzer range. If the analyzer again operates above 100% of its range, repeat the test using the next higher range. Continue to repeat the test until the analyzer always operates at less than 100% of its range.

(b) Drift verification. Gas analyzer drift verification is required for all gaseous exhaust constituents for which an emission standard applies. It is also required for CO2 even if there is no CO2 emission standard. It is not required for other gaseous exhaust constituents for which only a reporting requirement applies (such as CH4 and N2O).

(1) Verify drift using one of the following methods:

(i) For regulated exhaust constituents determined from the mass of a single component, perform drift verification based on the regulated constituent. For example, when NOX mass is determined with a dry sample measured with a CLD and the removed water is corrected based on measured CO2, CO, THC, and NOX concentrations, you must verify the calculated NOX value.

This document is only available to subscribers. Please log in or purchase access.